-Load cells -Pressure, torque and displacement transducers Features: -Front panel calibration and operation -Simulative or live load calibration -5 digit high brightness LED display
1.Analog signal Input range Min. 0.5mV/V, Max. 3.2mV/V 2.Non-linearity +0.02% of full scale 3.Internal resolution ±1/200,000 4.Temp. coefficient for Zero-0.25 V/ RTI Max. transducer or loadcell Span-within 0.01% reading/ 5.Excitation for pressure transducer or loadcell 6.Display Type LED, 7 segments red, 10mm high Active digit unit weight (00000~99990), Polarity indication "-" sign ("-"ranges are limited 30% of F.C) Display increments 1,2,5,10-selectable Decimal points Selectable to any points Status annunciators Zero alarm, auto-zero, relay point (L1~L4), tare 7.Sampling speed Appr. 30 times per second Relay output mode Mode 1: 4 relay outputs, 3 setpoints Mode 2: 4 relay outputs, 4 setpoints Mode 3: Relay by external signal Mode 4: Gross total, 4 relay, 3 setpoints Mode 5: Gross total, 4 relay, 4 setpoints 9.Operating Voltage AC ll0, 220V select by input power changeable holder 10.Power Consumption Appr. 18VA 11.Calibration Live load calibration by front panel push button 12.Hold mode Peak hold (only available on BS4220A), average hold 13.Interface 1. Analog Output Voltage output (DC 0-10V) 2. Analog Output Current output (DC 4-20mA) 3. Serial Output RS-232C output 14.Weight Appr. 2.3kg DC 12V, 300mA 8.Set point mode oC oC For: -Load cells -Pressure, torque and displacement transducers Features: -Front panel calibration and operation -Simulative or live load calibration -5 digit high brightness LED display The BS4200A series multifunction transducer Indicator/Controller are a high performance microprocessor based digital weight indicator and setpoint controller combined in one package. All setpoint settings are entered via the front panel keyboard, which in addition to the standard controls for weighing and setpoint entry, also includes controls for direct keyboard entry. The basic BS4200A comprises a loadcell amplifier and A/D converter, 4 user configurable output relays and excitation for up tomsix 350 ohm transducers in parallel. Alternatively, the BS4220A indicator incorporates total weight display and peak hold function. The basic BS4200A also accommodates a wide variety of boards providing 4-20A, 0-10V or RS-232C serial output. Built for rugged industrial use, the BS4200A features a steel enclosure and the unit will operate on 110VAC or 220VAC. Typical applications include weighing, batching, mixing, force, pressure, torque and displacement.
|